Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 111
Filtrar
1.
Diabetes Technol Ther ; 26(3): 190-197, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38444313

RESUMO

Aim: To assess the real-world performance of MiniMed™ 780G for Australians with type 1 diabetes (T1D) following advanced hybrid closed loop (AHCL) activation and to evaluate the effect of changing from MiniMed 670/770G to 780G. Methods: We analyzed deidentified Carelink™ continuous glucose monitoring (CGM) data from Australian users from January 2020 to December 2022, including the proportion attaining three major consensus targets: Glucose management indicator (GMI <7.0%), time in range (TIR 70-180 mg/dL >70%), and time below range (TBR 70 mg/dL <4%). Results: Comparing 670/770G users (n = 5676) for mean ± standard deviation 364 ± 244 days with 780G users (n = 3566) for 146 ± 145 days, the latter achieved a higher TIR (72.6% ± 10.6% vs. 67.3% ± 11.4%; P < 0.001), lower time above range (TAR) (25.5% ± 10.9% vs. 30.6% ± 11.7%; P < 0.001), and lower GMI (6.9% ± 0.4% vs. 7.2% ± 0.4%; P < 0.001) without compromising TBR (1.9% ± 1.8% vs. 2.0% ± 1.8%; P = 0.0015). Of 1051 670/770G users transitioning to 780G, TIR increased (70.0% ± 10.7% to 74.0% ± 10.2%; P < 0.001), TAR decreased (28.1% ± 10.9% to 24.0% ± 10.7%; P < 0.001), and TBR was unchanged. The percentage of users attaining all three CGM targets was higher in 780G users (50.1% vs. 29.5%; P < 0.001). CGM metrics were stable at 12 months post-transition. Conclusion: Real-world data from Australia shows that a higher proportion of MiniMed 780G users meet clinical targets for CGM consensus metrics compared to MiniMed 670/770G users and glucose control was sustained over 12 months.


Assuntos
População Australasiana , Automonitorização da Glicemia , Insulina , Humanos , Austrália , Glicemia , Insulina/uso terapêutico , Insulina Regular Humana
2.
Artigo em Inglês | MEDLINE | ID: mdl-38417017

RESUMO

Background: Carbohydrate counting (CC) and meal announcements, before eating, introduce a significant burden for individuals managing type 1 diabetes (T1D). An automated insulin delivery system with automatic bolusing that eliminates the need for CC and premeal bolusing (i.e., a hands-free closed-loop [HFCL] system) was assessed in a feasibility trial of adults with T1D. Methods: The system included the MiniMed™ 780G pump and a smartphone-paired smartwatch with the Klue application (Klue, Inc.) that detects eating and drinking gestures. A smartphone algorithm converted gestures into carb amounts that were transmitted to the pump for automatic bolusing. For 5 days, participants (N = 17, 18-75 years of age) used the system at home with meal announcements based on traditional CC, with the Klue application disabled (Home-stay phase). Thereafter, participants moved to a supervised hotel setting, where the Klue application was enabled for 5 days and meals were not announced (Hotel-stay phase). Participants consumed the same eight test meals (six solid and two liquid) of varying caloric and carb size at the same time and day of the week for both phases, and glycemic metrics were compared. Otherwise, there were no other meal restrictions. Results: The overall time in range (70-180 mg/dL) was 83.4% ± 7.0% and 80.6% ± 6.7% for the Home-stay and Hotel-stay, respectively (P = 0.08). The average time at <70 mg/dL was 3.1% and 3.0% (P = 0.9144), respectively, and the average time at >180 mg/dL was 13.5% and 16.3% (P = 0.1046), respectively. Postprandial glycemia following low-carb test meals was similar between the two phases. The system's ability to accommodate high-carb meals was somewhat limited. There were no episodes of severe hypoglycemia or diabetic ketoacidosis. Conclusion: Preliminary findings show that a HFCL system was safe and maintained overall glycemic control, similar to that observed with traditional CC and manual meal bolusing. By eliminating these daily T1D burdens, a HFCL system may improve quality of life for individuals with T1D. ClinicalTrials.gov number: NCT04964128.

3.
Diabetes Technol Ther ; 26(S3): 24-31, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38377317

RESUMO

Background: The MiniMed™ 780G system (MM780G) with Guardian™ 4 sensor includes a 100 mg/dL glucose target (GT) and automated insulin corrections up to every 5 min and was recently approved for use in the United States. In the present study, early real-world MM780G performance and the use of recommended system settings (100 mg/dL GT with an active insulin time of 2 h), by individuals with type 1 diabetes, were evaluated. Methods: CareLink™ personal data uploaded between the launch of the MM780G to August 22, 2023 were aggregated and underwent retrospective analysis (based on user consent) and if users had ≥10 days of continuous glucose monitoring (CGM) data. The 24-h day CGM metrics, including mean glucose, percentage of time spent in (%TIR), above (%TAR), and below (%TBR) target range (70-180 mg/dL), in addition to delivered insulin and closed-loop (CL) exits, were compared between an overall group (n = 7499) and individuals who used recommended settings (each, for >95% of the time). An analysis of the same metrics for MiniMed™ 770G system (MM770G) users (n = 3851) who upgraded to the MM780G was also conducted (paired t-test or Wilcoxon signed-rank test, P < 0.05 considered statistically significant). Results: For MM780G users, CGM use, and time in CL were >90% and all MM780G CGM metrics exceeded consensus-recommended goals. With recommended settings (22% of all users), mean %TIR and %TITR (70-140 mg/dL) were 81.4% and 56.4%, respectively. For individuals who upgraded from the MM770G, %TIR and %TITR increased from 73.2% to 78.3% and 45.8% to 52.6%, respectively, while %TAR reduced from 25.1% to 20.2% (P < 0.001, for all three). CL exits/week averaged <1, for all MM780G users. Conclusions: Early real-world MM780G use in the United States demonstrated a high percentage of time in range with low time above and below range. These outcomes are similar to those observed for real-world MM780G use in other countries.


Assuntos
Glicemia , Diabetes Mellitus Tipo 1 , Estados Unidos , Humanos , Automonitorização da Glicemia , Estudos Retrospectivos , Insulina/uso terapêutico , Insulina Regular Humana , Diabetes Mellitus Tipo 1/tratamento farmacológico , Glucose , Hipoglicemiantes/uso terapêutico , Sistemas de Infusão de Insulina
4.
Diabetes Technol Ther ; 26(S3): 66-75, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38377319

RESUMO

Intensive therapy with exogenous insulin is the treatment of choice for individuals living with type 1 diabetes (T1D) and some with type 2 diabetes, alongside regular glucose monitoring. The development of systems allowing (semi-)automated insulin delivery (AID), by connecting glucose sensors with insulin pumps and algorithms, has revolutionized insulin therapy. Indeed, AID systems have demonstrated a proven impact on overall glucose control, as indicated by effects on glycated hemoglobin (HbA1c), risk of severe hypoglycemia, and quality of life measures. An alternative endpoint for glucose control that has arisen from the use of sensor-based continuous glucose monitoring is the time in range (TIR) measure, which offers an indication of overall glucose control, while adding information on the quality of control with regard to blood glucose level stability. A review of literature on the health-economic value of AID systems was conducted, with a focus placed on the growing place of TIR as an endpoint in studies involving AID systems. Results showed that the majority of economic evaluations of AID systems focused on individuals with T1D and found AID systems to be cost-effective. Most studies incorporated HbA1c, rather than TIR, as a clinical endpoint to determine treatment effects on glucose control and subsequent quality-adjusted life year (QALY) gains. Likely reasons for the choice of HbA1c as the chosen endpoint is the use of this metric in most validated and established economic models, as well as the limited publicly available evidence on appropriate methodologies for TIR data incorporation within conventional economic evaluations. Future studies could include the novel TIR metric in health-economic evaluations as an additional measure of treatment effects and subsequent QALY gains, to facilitate a holistic representation of the impact of AID systems on glycemic control. This would provide decision makers with robust evidence to inform future recommendations for health care interventions.


Assuntos
Diabetes Mellitus Tipo 1 , Diabetes Mellitus Tipo 2 , Humanos , Diabetes Mellitus Tipo 1/tratamento farmacológico , Diabetes Mellitus Tipo 2/tratamento farmacológico , Hipoglicemiantes , Hemoglobinas Glicadas , Glicemia/metabolismo , Automonitorização da Glicemia/métodos , Qualidade de Vida , Insulina , Sistemas de Infusão de Insulina , Insulina Regular Humana/uso terapêutico
5.
Diabetes Technol Ther ; 26(S3): 38-44, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38377323

RESUMO

The mean absolute relative difference (MARD) is a numerical metric that has been adopted by the diabetes technology community as the main indicator that describes the accuracy of a glucose sensor at a single point in time. The appropriateness of this adoption is questionable because there is limited evidence that MARD has meaningful clinical relevance in the current era of sensor technology. The calculation may be simple, but evaluation of MARD can be very complex because it is substantially impacted by the design of the data collection in an accuracy study. Factors that can influence the overall MARD include participant demographics such as type of diabetes and age, site of sensor wear, and the percentage of collected values in each glycemic range during the study that is, in turn, a function of the study design. MARD is only one of several important statistical metrics such as bias and precision that are relevant to assessing accuracy of a sensor. Furthermore, these analytic metrics convey little information about the safety and effectiveness of sensor use with an automated insulin delivery system or a standalone device. There are no clinical studies in people with diabetes (PWD) proving that MARD can accurately differentiate between a safe and unsafe sensor or between a more and less clinically effective sensor. Moreover, there are alternatives to MARD that can do this in a clinically meaningful way, which include error grid analyses and clinical studies in PWD. This review attempts to demythologize the status of MARD for the diabetes community in an effort to shift the focus from MARD to using clinically relevant assessments.


Assuntos
Diabetes Mellitus Tipo 1 , Humanos , Diabetes Mellitus Tipo 1/tratamento farmacológico , Glicemia , Automonitorização da Glicemia , 60431 , Reprodutibilidade dos Testes
6.
Diabetes Technol Ther ; 26(S3): 17-23, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38377324

RESUMO

The MiniMed™ 780G system (780G) received Conformité Européenne mark in June 2020 and was, recently, approved by the U.S. Food and Drug Administration (April 2023). Clinical trials and real-world analyses have demonstrated MiniMed™ 780G system safety and effectiveness and that glycemic outcomes (i.e., time in range) improve with recommended settings use. In this publication, we will explain the iterative development of the 780G algorithm and how this technology has simplified diabetes management.


Assuntos
Diabetes Mellitus Tipo 1 , Hipoglicemiantes , Humanos , Hipoglicemiantes/uso terapêutico , Glicemia/análise , Diabetes Mellitus Tipo 1/tratamento farmacológico , Insulina/uso terapêutico , Sistemas de Infusão de Insulina , Automonitorização da Glicemia , Algoritmos
7.
Diabetes Technol Ther ; 26(1): 70-75, 2024 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-37955697

RESUMO

Continuous glucose monitoring (CGM) is proposed as an alternative for glycemic assessment in peritoneal dialysis, but volume overload and anemia may affect sensor accuracy. This is an exploratory analysis of a study of Guardian Connect™ with Guardian Sensor™ 3 in 30 participants with diabetes on continuous ambulatory peritoneal dialysis (CAPD) (age [mean ± standard deviation] 64.7 ± 5.6 years, 23 men, body mass index [BMI] 25.4 ± 3.9 kg/m2, blood hemoglobin [Hb] 10.7 ± 1.3 g/dL). The mean absolute relative difference (MARD) was calculated between paired sensor and YSI 2300 STAT venous glucose readings (n = 941) during an 8-h in-clinic session with glucose challenge. Body composition was evaluated using bioimpedance. The overall MARD was 10.4% (95% confidence interval 9.6-11.7). There were no correlations between BMI, extracellular water, relative hydration index, and lean or fat mass with MARD. No correlations were observed between MARD and Hb (r = 0.016, P > 0.05). In summary, this real-time CGM demonstrated good accuracy in CAPD with minimal influence from body composition and anemia.


Assuntos
Anemia , Diabetes Mellitus Tipo 1 , Diálise Peritoneal Ambulatorial Contínua , Masculino , Humanos , Pessoa de Meia-Idade , Idoso , Glicemia , Automonitorização da Glicemia , Reprodutibilidade dos Testes , Anemia/etiologia , Composição Corporal
8.
Diabetes Technol Ther ; 26(1): 33-39, 2024 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-37855818

RESUMO

Objective: Connected insulin pens are creating opportunities for the millions of individuals with diabetes using multiple daily injections (MDI) therapy across the globe. Continuous glucose monitoring (CGM) data from connected insulin pens are revealing gaps and opportunities to significantly improve care for this population. In this article, we report real-world findings of the InPen™ smart insulin pen paired with CGM (InPen system), used by persons with type 1 diabetes (T1D) and type 2 diabetes (T2D). Methods: A retrospective cohort analysis was conducted with the real-world data collected from the InPen system of individuals (N = 3793 with T1D, N = 552 with T2D, and N = 808 unidentified) who used the system from January 01, 2020, to December 31, 2021. Diabetes management (e.g., missed and mistimed insulin dosing, mismatched food intake, and correction dose delivery) and glycemic outcomes were assessed. Results: In the overall and T1D populations, a dosing frequency of ≥3 doses per day and a missed dose frequency of <20% was associated with improved glycemia. In adults with T2D, missing <20% of doses was the significant factor determining improved glycemia. Conclusion: This analysis, integrating data from a smart insulin pen and CGM, provides insights into the impact of dosing behavior on glycemic outcomes and informs counseling strategies for the diabetes care team, through technologically advanced insulin management for those using MDI therapy.


Assuntos
Diabetes Mellitus Tipo 1 , Diabetes Mellitus Tipo 2 , Adulto , Humanos , Insulina/uso terapêutico , Diabetes Mellitus Tipo 1/tratamento farmacológico , Diabetes Mellitus Tipo 2/tratamento farmacológico , Hipoglicemiantes/uso terapêutico , Automonitorização da Glicemia , Estudos Retrospectivos , Glicemia , Insulina Regular Humana/uso terapêutico
9.
Diabetes Technol Ther ; 25(11): 755-764, 2023 11.
Artigo em Inglês | MEDLINE | ID: mdl-37782145

RESUMO

Background: During MiniMed™ advanced hybrid closed-loop (AHCL) use by adolescents and adults in the pivotal trial, glycated hemoglobin (A1C) was significantly reduced, time spent in range (TIR) was significantly increased, and there were no episodes of severe hypoglycemia or diabetic ketoacidosis (DKA). The present study investigated the same primary safety and effectiveness endpoints during AHCL use by a younger cohort with type 1 diabetes (T1D). Methods: An intention-to-treat population (N = 160, aged 7-17 years) with T1D was enrolled in a single-arm study at 13 investigational centers. There was a run-in period (∼25 days) using HCL or sensor-augmented pump with/without predictive low-glucose management, followed by a 3-month study period with AHCL activated at two glucose targets (GTs; 100 and 120 mg/dL) for ∼45 days each. The mean ± standard deviation values of A1C, TIR, mean sensor glucose (SG), coefficient of variation (CV) of SG, time at SG ranges, and insulin delivered between run-in and study were analyzed (Wilcoxon signed-rank test or t-test). Results: Compared with baseline, AHCL use was associated with reduced A1C from 7.9 ± 0.9% (N = 160) to 7.4 ± 0.7% (N = 136) (P < 0.001) and overall TIR increased from the run-in 59.4 ± 11.8% to 70.3 ± 6.5% by end of study (P < 0.001), without change in CV, time spent below range (TBR) <70 mg/dL, or TBR <54 mg/dL. Relative to longer active insulin time (AIT) settings (N = 52), an AIT of 2 h (N = 19) with the 100 mg/dL GT increased mean TIR to 73.4%, reduced TBR <70 mg/dL from 3.5% to 2.2%, and reduced time spent above range (TAR) >180 mg/dL from 28.7% to 24.4%. During AHCL use, there was no severe hypoglycemia or DKA. Conclusions: In children and adolescents with T1D, MiniMed AHCL system use was safe, A1C was lower, and TIR was increased. The lowest GT and shortest AIT were associated with the highest TIR and lowest TBR and TAR, all of which met consensus-recommended glycemic targets. ClinicalTrials.gov ID: NCT03959423.


Assuntos
Diabetes Mellitus Tipo 1 , Cetoacidose Diabética , Hipoglicemia , Adolescente , Adulto , Criança , Humanos , Glicemia , Automonitorização da Glicemia , Diabetes Mellitus Tipo 1/tratamento farmacológico , Diabetes Mellitus Tipo 1/complicações , Cetoacidose Diabética/etiologia , Glucose , Hemoglobinas Glicadas , Hipoglicemia/induzido quimicamente , Hipoglicemia/prevenção & controle , Hipoglicemia/complicações , Hipoglicemiantes/uso terapêutico , Insulina/uso terapêutico , Sistemas de Infusão de Insulina , Resultado do Tratamento
10.
J Diabetes Sci Technol ; : 19322968231202803, 2023 Sep 24.
Artigo em Inglês | MEDLINE | ID: mdl-37743727

RESUMO

BACKGROUND: Recent evidence shows that it may be safe to estimate bolus sizes based on continuous glucose monitoring (CGM) rather than blood glucose (BG) values using glycemic trend-adjusted bolus calculators. Users may already be doing this in the real world, though it is unclear whether this is safe or effective for calculators not employing trend adjustment. METHODS: We assessed real-world data from a smart multiple daily injections (MDIs) device users with a CGM system, hypothesizing that four-hour post-bolus outcomes using CGM values are not inferior to those using BG values. Our data set included 184 users and spanned 18 months with 79 000 bolus observations. We tested differences using logistic regression predicting CGM or BG value usage based on outcomes and confirmed initial results using a mixed model regression accounting for within-subject correlations. RESULTS: Comparing four-hour outcomes for bolus events using CGM and BG values revealed no differences using our initial approach (P > .183). This finding was confirmed by our mixed model regression approach in all cases (P > .199), except for times below range outcomes. Higher times below range were predictive of lower odds of CGM-based bolus calculations (OR = 0.987, P < .0001 and OR = 0.987, P = .0276, for time below 70 and 54 mg/dL, respectively). CONCLUSIONS: We found no differences in four-hour post-bolus glycemic outcomes when using CGM or BG except for time below range, which showed evidence of being lower for CGM. Though preliminary, our results confirm prior findings showing non-inferiority of using CGM values for bolus calculation compared with BG usage in the real world.

11.
Diabetes Technol Ther ; 25(10): 718-725, 2023 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-37578804

RESUMO

Background: While evidence supports glycemic control benefits for individuals with type 1 diabetes mellitus (T1DM) using hybrid closed-loop (HCL) systems, HCL automated insulin delivery therapy in China has not been assessed. This study evaluated safety events and effectiveness during HCL system use by Chinese adolescents and adults with T1DM. Methods: Sixty-two participants (n = 12 adolescents with a mean ± standard deviation [SD] of 15.5 ± 1.1 years and n = 50 adults [mean ± SD of 37.6 ± 11.1 years]) with T1DM and baseline A1C of 7.1% ± 1.0% underwent a run-in period (∼2 weeks) using open-loop Manual Mode (sensor-augmented pump) insulin delivery with the MiniMed™ 770G system with the Guardian™ Sensor (3) glucose sensor, followed by a study period (4 weeks) with HCL Auto Mode enabled. Analyses compared continuous glucose monitoring data and insulin delivered during the run-in versus study period (Wilcoxon signed-rank test or t-test). Safety events included rates of severe hypoglycemia and diabetic ketoacidosis (DKA). Results: Compared to baseline run-in, overall Auto Mode use increased time in range (TIR, 70-180 mg/dL) from 75.3% to 80.9% (P < 0.001) and reduced time below range (TBR, <70 mg/dL) from 4.7% to 2.2% (P < 0.001). Subgroup analysis demonstrated that participants (n = 29) with baseline A1C <7.0% had TBR that reduced from 5.6% to 2.0%, while participants (n = 21) with baseline A1C ≥7.5% had time above range (TAR, >180 mg/dL) that reduced from 31.6% to 20.8%. Auto Mode use also increased the percentage achieving combined recommendations for time at sensor glucose ranges (i.e., TIR of >70%, TBR of <4% and TAR of <25%) from 24.2% at baseline to 77.4% at study end. Total daily insulin dose reduced from 42.8 ± 19.8 to 40.7 ± 18.9 U (P = 0.013). There were no severe hypoglycemic, DKA, or serious adverse events. Conclusions: Chinese adolescents and adults, some of whom met target A1C at baseline, safely achieved significantly improved glycemia with 1 month of MiniMed 770G system use when compared to open-loop insulin delivery. ClinicalTrials.gov ID: NCT04663295.

12.
Diabetes Obes Metab ; 25(9): 2457-2463, 2023 09.
Artigo em Inglês | MEDLINE | ID: mdl-37353345

RESUMO

AIM: To investigate the association between a new composite metric, glycaemia risk index (GRI), and incident diabetic retinopathy (DR). METHODS: A total of 1204 adults with type 2 diabetes without DR at baseline were included between 2005 and 2019 from a single centre in Shanghai, China. GRI was obtained from continuous glucose monitoring data at baseline. Cox proportion hazard regression analysis was used to assess the association between GRI and the risk of incident DR. RESULTS: During a median follow-up of 8.4 years, 301 patients developed DR. The multivariable-adjusted hazard ratios (HRs) for incident DR across ascending GRI quartiles (≤14 [reference], 15 ~ 28, 29 ~ 47 and > 47) were 1.00, 1.05 (95% CI 0.74-1.48), 1.33 (95% confidence interval [CI] 0.96-1.84) and 1.53 (95% CI 1.11-2.11), respectively. For each 1-SD increase in GRI, the risk of DR was increased by 20% (HR 1.20, 95% CI 1.07-1.33) after adjustment for confounders. CONCLUSIONS: In patients with type 2 diabetes, higher GRI is associated with an increased risk of incident DR. GRI has the potential to be a valuable clinical measure, which needs to be further explored in future studies.


Assuntos
Diabetes Mellitus Tipo 2 , Retinopatia Diabética , Adulto , Humanos , Diabetes Mellitus Tipo 2/complicações , Diabetes Mellitus Tipo 2/epidemiologia , Estudos de Coortes , Retinopatia Diabética/etiologia , Retinopatia Diabética/complicações , Fatores de Risco , Automonitorização da Glicemia , Glicemia , China/epidemiologia
13.
Diabetes Technol Ther ; 25(9): 652-658, 2023 09.
Artigo em Inglês | MEDLINE | ID: mdl-37252734

RESUMO

Background: Safety and significant improvement in overall glycated hemoglobin (A1C) and percentage of time spent in (TIR), below (TBR), and above (TAR) glucose range were demonstrated in the pivotal trial of adolescents and adults using the MiniMed™ advanced hybrid closed-loop (AHCL) system with the adjunctive, calibration-required Guardian™ Sensor 3. The present study evaluated early outcomes of continued access study (CAS) participants who transitioned from the pivotal trial investigational system to the approved MiniMed™ 780G system with the non-adjunctive, calibration-free Guardian™ 4 Sensor (MM780G+G4S). Study data were presented alongside those of real-world MM780G+G4S users from Europe, the Middle East, and Africa. Methods: The CAS participants (N = 109, aged 7-17 years and N = 67, aged >17 years) used the MM780G+G4S for 3 months and data of real-world MM780G+G4S system users (N = 10,204 aged ≤15 years and N = 26,099 aged >15 years) were uploaded from September 22, 2021 to December 02, 2022. At least 10 days of real-world continuous glucose monitoring (CGM) data were required for analyses. Glycemic metrics, delivered insulin and system use/interactions underwent descriptive analyses. Results: Time in AHCL and CGM use were >90% for all groups. AHCL exits averaged 0.1/day and there were few blood glucose measurements (BGMs) (0.8/day-1.0/day). Adults in both cohorts met most consensus recommendations for glycemic targets. Pediatric groups met recommendations for %TIR and %TBR, although not those for mean glucose variability and %TAR, possibly due to low use of recommended glucose target (100 mg/dL) and active insulin time (2 h) settings (28.4% in the CAS cohort and 9.4% in the real-world cohort). The CAS pediatric and adult A1C were 7.2% ± 0.7% and 6.8% ± 0.7%, respectively, and there were no serious adverse events. Conclusions: Early clinical use of the MM780G+G4S was safe and involved minimal BGMs and AHCL exits. Consistent with real-world pediatric and adult use, outcomes were associated with achievement of recommended glycemic targets. Clinical Trial Registration number: NCT03959423.


Assuntos
Glicemia , Diabetes Mellitus Tipo 1 , Adolescente , Adulto , Criança , Humanos , Automonitorização da Glicemia , Diabetes Mellitus Tipo 1/tratamento farmacológico , Glucose , Hemoglobinas Glicadas , Hipoglicemiantes/uso terapêutico , Insulina/uso terapêutico , Sistemas de Infusão de Insulina
14.
J Diabetes Sci Technol ; : 19322968231161320, 2023 Mar 22.
Artigo em Inglês | MEDLINE | ID: mdl-36949671

RESUMO

BACKGROUND: This analysis reports the findings from a predefined exploratory cohort (cohort B) from the ADAPT (ADvanced Hybrid Closed Loop Study in Adult Population with Type 1 Diabetes) study. Adults with type 1 diabetes (T1D) with suboptimal glucose control were randomly allocated to an advanced hybrid closed-loop (AHCL) system or multiple daily injections of insulin (MDI) plus real-time continuous glucose monitoring (RT-CGM). METHODS: In this prospective, multicenter, exploratory, open-label, randomized controlled trial, 13 participants using MDI + RT-CGM and with HbA1c ≥8.0% were randomized to switch to AHCL (n = 8) or continue with MDI + RT-CGM (n = 5) for six months. Prespecified endpoints included the between-group difference in mean change from baseline in HbA1c, CGM-derived measures of glycemic control, and safety. RESULTS: The mean HbA1c level decreased by 1.70 percentage points in the AHCL group versus a 0.60 percentage point decrease in the MDI + RT-CGM group, with a model-based treatment effect of -1.08 percentage points (95% confidence interval [CI] = -2.17 to 0.00 percentage points; P = .0508) in favor of AHCL. The percentage of time spent with sensor glucose levels between 70 and 180 mg/dL in the study phase was 73.6% in the AHCL group and 46.4% in the MDI + RT-CGM group; model-based between-group difference of 28.8 percentage points (95% CI = 12.3 to 45.3 percentage points; P = .0035). No diabetic ketoacidosis or severe hypoglycemia occurred in either group. CONCLUSIONS: In people with T1D with HbA1c ≥8.0%, the use of AHCL resulted in improved glycemic control relative to MDI + RT-CGM. The scale of improvement suggests that AHCL should be considered as an option for people not achieving good glycemic control on MDI + RT-CGM.

15.
J Manag Care Spec Pharm ; 29(3): 285-292, 2023 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-36692907

RESUMO

BACKGROUND: Severe hypoglycemia is a significant barrier to optimizing insulin therapy in both type 1 and type 2 diabetes and places a burden on the US health care system because of the high costs of hypoglycemia-related health care utilization. OBJECTIVE: To compare the frequency of sensor-detected severe hypoglycemic events (SHEs) among a population of continuous glucose monitoring (CGM) users on insulin therapy after initiation of the InPen smart insulin pen (SIP) system and to estimate the potential hypoglycemia-related medical cost savings across a population of SIP users. METHODS: SIP users of all ages with type 1 or type 2 diabetes were required to have at least 90 days of SIP use with a connected CGM device. The last 14 days of sensor glucose (SG) data within the 30-day period prior to the start of SIP use ("pre-SIP") and the last 14 days of SG data, along with the requirement of at least 1 bolus entry per day within the 61- to 90-day period after SIP start ("post-SIP"), were analyzed. Sensor-detected SHEs (defined as ≥10 minutes of consecutive SG readings at <54 mg/dL) were determined. Once factored, the expected medical intervention rates and associated costs were calculated. Intervention rates and costs were obtained from the literature. RESULTS: There were 1,681 SIP + CGM users from March 1, 2018, to April 30, 2021. The mean number of sensor-detected SHEs per week declined from 0.67 in the pre-SIP period to 0.58 in the post-SIP period (P = 0.008), which represented a 13% reduction. Assuming a range of 5%-25% of all sensor-detected SHEs resulted in a clinical event, the estimated cost reduction associated with reduced SHEs was $12-$59 and $110-$551 per SIP user per month and per year, respectively. For those aged at least 65 years, there were 166 SIP+CGM users and the reduction in the mean number of sensor-detected SHEs per week between the pre-SIP and post-SIP periods was 31%. CONCLUSIONS: Use of the SIP system with a connected CGM is associated with reduced sensor-detected severe hypoglycemia, which may result in significant cost savings. DISCLOSURES: Albert Chien, Glen Im, Kael Wherry, Janice MacLeod, and Robert A Vigersky are employees of Medtronic; Sneha Thanasekaran and Angela Gaetano were affiliated with Medtronic while doing this research. The submitted work did not involve study subject recruitment, enrollment, or participation in a trial and did not fall under human subject protection requirements (per the Department of Health and Human Services CFR Part 46) necessitating Internal Review Board approval or exemption.


Assuntos
Diabetes Mellitus Tipo 1 , Diabetes Mellitus Tipo 2 , Hipoglicemia , Humanos , Glicemia , Automonitorização da Glicemia/métodos , Redução de Custos , Diabetes Mellitus Tipo 1/tratamento farmacológico , Diabetes Mellitus Tipo 2/tratamento farmacológico , Glucose/uso terapêutico , Hipoglicemia/epidemiologia , Hipoglicemiantes/uso terapêutico , Insulina/uso terapêutico , Estados Unidos
16.
J Clin Endocrinol Metab ; 108(7): e388-e395, 2023 06 16.
Artigo em Inglês | MEDLINE | ID: mdl-36715293

RESUMO

CONTEXT: Racial/ethnic inequities have been observed in diabetes care. OBJECTIVE: To measure changes in prevalence of continuous glucose monitoring (CGM) and insulin pump therapy among Medicare Advantage beneficiaries with type 1 diabetes by race/ethnicity and to determine the impact of socioeconomic factors on racial/ethnic inequities. DESIGN: The prevalence of CGM and pump use was assessed by race/ethnicity for Medicare Advantage beneficiaries annually from 2017 through 2020. Models predicting technology use by year, race/ethnicity, age, sex, endocrinology visits, and measures of socioeconomic status were fit. SETTING: Community. PATIENTS OR OTHER PARTICIPANTS: Beneficiaries with type 1 diabetes and 2 or more claims with a diabetes diagnosis in the coverage year. INTERVENTION(S): Insulin pump or CGM therapy. MAIN OUTCOME MEASURE(S): Use of diabetes technology by racial/ethnic group. RESULTS: Technology use increased from 2017 through 2020 in all racial/ethnic groups. The absolute difference in use between White and Black beneficiaries from 2017 to 2020 remained stable for insulin pumps (10.7% to 10.8%) and increased for CGM (2.6% to 11.1%). The differences in pump use from 2017 to 2020 narrowed between White and Hispanic beneficiaries (12.3% to 11.4%) and White and Asian beneficiaries (9.7% to 6.6%), whereas the opposite occurred for CGM use (3.0% to 15.5% for White vs Hispanic beneficiaries; 1.5% to 8.0% for White vs Asian beneficiaries). Racial/ethnic inequities persisted (P < .0001) after adjusting for other characteristics. CONCLUSIONS: Differences in diabetes technology use between racial/ethnic groups often persisted from 2017 through 2020 and could not be explained by demographics, socioeconomic status, or endocrinology visits.


Assuntos
Diabetes Mellitus Tipo 1 , Insulinas , Medicare Part C , Humanos , Idoso , Estados Unidos/epidemiologia , Diabetes Mellitus Tipo 1/tratamento farmacológico , Automonitorização da Glicemia , Glicemia
17.
Ann Biomed Eng ; 51(1): 211-224, 2023 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-36125605

RESUMO

In silico simulation studies using 5807 virtual patients with insulin dependent diabetes have been conducted to estimate the risk and efficacy with the closed-loop 780G pump when switching between Medtronic Guardian Sensor 3 (GS3) and Medtronic Guardian 4 Sensor (G4S), next generation calibration free glucose sensor. To demonstrate by utilizing a case study that captures the merits of in silico studies with single hormone insulin dependent virtual patients that include variability in pharmacokinetics/pharmacodynamics, age, gender, insulin sensitivity and BMIs. Also, to show that in silico studies can uniquely isolate the effect of a single variable on clinical outcomes. Simulation studies results were compared to clinical and commercial data and were separated by age groups and pump settings. The commercial data, the clinical study data and the simulation studies predicted that switching between GS3 to G4S will introduce a change in glucose average, percentage time between 70 and 180 mg/dL, and percentage time below 70 mg/dL of: 5.2, 3.4, and 3.1 mg/dL, - 1.1, 0.2, and - 1.1%, and - 0.6, - 1.0, and - 0.3%, respectively. We demonstrated that our simulation studies were able to predict the difference in glycemic outcomes when switching between different sensors in real world setting, better than a small clinical controlled study. As predicted, switching between GS3 and G4S sensors with the 780G system does not introduce clinical risk and maintain the clinical outcomes of the sensor. We demonstrated the ability of insulin dependent diabetes virtual patients to predict clinical outcomes and to augment or even replace some small clinical studies.


Assuntos
Diabetes Mellitus Tipo 1 , Hipoglicemiantes , Humanos , Hipoglicemiantes/uso terapêutico , Glicemia , Glucose , Insulina
18.
J Diabetes Sci Technol ; 17(4): 951-958, 2023 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-35414272

RESUMO

BACKGROUND: Glycemic outcomes during real-world hybrid closed-loop (HCL) system use by individuals with type 1 diabetes, in the United States, were retrospectively analyzed. METHODS: Hybrid closed-loop system data voluntarily uploaded to Carelink™ personal software from March 2017 to November 2020 by individuals (aged ≥7 years) using the MiniMed™ 670G system and having ≥10 days of continuous glucose monitoring data after initiating Auto Mode were assessed. Glycemic outcomes including the mean glucose management indicator (GMI), sensor glucose (SG), percentage of time spent in (TIR), below (TBR), and above (TAR) target range (70-180 mg/dL) were analyzed. Outcomes were also analyzed in a subgroup of users per baseline GMI of <7% versus >8%. RESULTS: The overall cohort (N = 123 355 users, with a mean of 87.9% of time in Auto Mode) had a GMI of 7.0% ± 0.4%, TIR of 70.4% ± 11.2%, TBR <70 mg/dL of 2.2% ± 2.1% and TAR>180 mg/dL of 27.5% ± 11.6%, post-Auto Mode initiation. Compared with pre-Auto Mode initiation, users (N = 52 941, 88.6% of time in Auto Mode) had a GMI that decreased from 7.3% ± 0.6% to 7.1% ± 0.5% (P < .001), TIR that increased from 61.5% ± 15.1% to 68.1% ± 11.9% (P < .001), TAR>180 mg/dL that decreased from 36.3% ± 15.7% to 29.8% ± 12.2% (P < .001) and TBR<70 mg/dL that decreased from 2.11 ± 2.4 to 2.07% ± 2.25% (P = .002). While all metrics statistically improved for the baseline GMI >8.0% group, the baseline GMI <7.0% group had unchanged TIR (77.4% ± 7.4% to 77.5% ± 8.0%, P = .456) and TAR>180 mg/dL that increased (19.2 ± 6.7 to 19.6 ± 7.9%, p < 0.001). CONCLUSION: Real-world HCL system use in the U.S. demonstrated overall glycemic control that trended similarly with the system pivotal trial outcomes and previous real-world system use analyses.


Assuntos
Diabetes Mellitus Tipo 1 , Humanos , Estados Unidos , Diabetes Mellitus Tipo 1/tratamento farmacológico , Hipoglicemiantes/uso terapêutico , Glicemia , Automonitorização da Glicemia , Insulina/uso terapêutico , Estudos Retrospectivos , Sistemas de Infusão de Insulina , Glucose
19.
Diabetes Technol Ther ; 25(2): 108-115, 2023 02.
Artigo em Inglês | MEDLINE | ID: mdl-36315189

RESUMO

Aim: To assess relationships between continuous glucose monitoring (CGM) time in range (TIR), 70-180 mg/dL, time below range (TBR), <70 mg/dL, time above range (TAR), >180 mg/dL, and glucose coefficient of variation (CV) in relation to currently recommended clinical CGM targets for older people, which recommend reduced TIR and TBR targets relative to the general type 1 diabetes population. Methods: We conducted a post hoc analysis using the JDRF Australia Adult Hybrid Closed Loop trial database examining correlations in 120 adults with type 1 diabetes of 3 weeks masked CGM (Guardian Sensor 3; Medtronic) metrics (n = 61 on multiple daily injections, 59 on non-CGM augmented pumps) using manual insulin dosing at baseline and at 26-weeks, with 50% randomized to automated insulin dosing (AID). Results: Correlations between baseline TIR and TAR were strong (r = -0.966; P < 0.0001), weak for TBR (r = 0.363; P < 0.0001), and glucose CV (r = 0.037; P = 0.687) while moderate between CV and TBR (r = 0.726; P < 0.0001). Associations were similar for participants aged >60 years (n = 15) versus younger subjects. Correlations of changes in (Δ) TIR with ΔTAR over 26 weeks were strong (r = -0.945; P < 0.001) and correlations for ΔTBR were weak (r = 0.025; P = 0.802). ΔCV did not significantly correlate with ΔTAR (r = -0.064; P = 0.526) but did with ΔTBR (r = 0.770; P = <0.001). Conclusions: Changes in TIR are not associated with changes in TBR. Thus, we recommend that for older AID users whilst TBR targets should be prioritized to reduce hypoglycemia-related risk, TBR should be addressed independently of TIR. Clinical Trial Registratrion number: (ACTRN12617000520336).


Assuntos
Diabetes Mellitus Tipo 1 , Adulto , Humanos , Idoso , Diabetes Mellitus Tipo 1/tratamento farmacológico , Glicemia , Hipoglicemiantes/uso terapêutico , Automonitorização da Glicemia , Insulina/uso terapêutico
20.
Diabetes Obes Metab ; 25(1): 110-120, 2023 01.
Artigo em Inglês | MEDLINE | ID: mdl-36053813

RESUMO

AIM: To evaluate the efficacy of intermittent short-term use of a real-time continuous glucose monitoring (RT-CGM) system in non-insulin-treated patients with type 2 diabetes (T2D) uncontrolled with oral antidiabetic drugs (OADs). MATERIALS AND METHODS: In this multicentre, randomized prospective study, 61 participants were randomly assigned to treatment group 1 (one session of RT-CGM), treatment group 2 (two sessions of RT-CGM with a 3-month interval between sessions) and a control group. All participants used blinded continuous glucose monitoring for up to 6 days with education before randomization, and RT-CGM was additionally applied for 1 week in the intervention groups. The primary outcome was change in HbA1c at 6 months. RESULTS: Among 61 participants, 48 subjects completed the study (baseline HbA1c 8.2% ± 0.5%). At 3 months, a significant HbA1c reduction was observed in treatment group 1 (adjusted difference = -0.60%, P = .044) and treatment group 2 (adjusted difference = -0.64%, P = .014) compared with the control group. However, at 6 months, only treatment group 2 achieved a significant HbA1c reduction (adjusted difference = -0.68%, P = .018). Especially in the treatment groups, patients performing self-monitoring of blood glucose (SMBG) at least 1.5 times/day showed a significant HbA1c improvement, at both 3 and 6 months, but those performing SMBG less than 1.5 times/day showed no significant improvement. CONCLUSIONS: In non-insulin-treated patients with T2D uncontrolled with OADs, intermittent short-term use of RT-CGM was an effective method for glucose control, especially in those performing SMBG frequently.


Assuntos
Diabetes Mellitus Tipo 2 , Humanos , Diabetes Mellitus Tipo 2/tratamento farmacológico , Automonitorização da Glicemia , Glicemia , Estudos Prospectivos